GENERATION OF CT SCANNERS
A variety of CT geometries have been developed to acquire the
X-ray transmission data for image reconstruction. These geometries
are commonly called generations. The main objective of different
generation is (i) scanning time reduction and (ii) simplification of
mechanical motion.
FIRST GENERATION
The first generation CT scanner is a rotate/translate, pencil beam system.
It had two X-ray detectors and used parallel ray geometry with NaI detector . It is translated linearly to acquire 160 rays across
a 24 cm FOV and rotated between translations to acquire 180 projections
at 1° interval. It took about 4.5 minutes per scan with 1.5 minutes
to reconstruct a slice with a linear measurements of 28,800 rays (160
× 180).
There is a large change in signal due to increased X-ray flux outside
of head and hence patient’s head is pressed into a flexible membrane
surrounded by a water bath. The NaI detector signal decayed slowly,
affecting measurements. The advantage of the system is the efficient
scatter reduction, best of all scanner generations. The disadvantage
includes (i) water bath acted as a bolus and (ii) afterglow of NaI.
SECOND GENERATION
The second generation CT scanner is also rotate/translate system
(11.11B), with narrow beam geometry (10°). Linear array of 30 detectors
were used to acquire more data, to improve image quality (600 rays
× 540 views = 3,24000). These scanners provided larger rotational
increments and faster scans. The shortest scan time was 18 seconds
per slice. Narrow fan beam allows more scattered radiation to be detected.
THIRD GENERATION
The third generation scanner is a rotate/rotate system with wide
beam geometry (Fig.11.12A). The number of detectors has increased
substantially (> 800 detectors) and the angle of fan beam is increased
to cover entire patient. It eliminated the need for translational motion.
The X-ray tube and detector array are mechanically joined androtate together.
Newer systems have scan times of the order of
< 0.5 second.
The 3rd generation scanners lead to a situation in which each detector
is responsible for the data corresponding to a ring in the image. Any
drift in the signal levels of the detectors over time affects the mt values
that are back projected to produce the CT image, causing ring artifacts.
FOURTH GENERATION
The fourth generation scanners are designed to overcome the problem
of ring artifacts. It has a stationary ring of about 4,800 detectors, and
the X-ray tube has to move inside this detector. Since it is rotated
continuously, very fast scan time is possible. Geometrical misalignment
between detector ring radius and X-ray beam origin may be possible.
It has inter scan delay times, since the X-ray tube had to return to
its starting position (home).
Third generation fan beam geometry has the X-ray tube as the
apex of the fan. In the 4th generation, the individual detector is the
apex . Though the X-ray tube forms the fan beam, data
are processed for fan beam reconstruction with each detector as the
vertex of a fan. The rays acquired by each detector are fanned out
to different positions of the X-ray source. In the 3rd generation, It
and Io are measured at the center and at the edge of the detector
array. Hence, the gain of the reference detector and the individual
detector may not be equal. In the 4th generation, each detector has
its own reference detector, and hence the gain of the reference and
individual detector is equal.
FIFTH GENERATION
The fifth generation scanner is a stationary/stationary system, developed
specifically for cardiac tomography imaging. No conventional X-ray tube
is used, instead large arc of tungsten (210°) encircles patient and lies
directly opposite to the detector ring. It uses an electron gun that deflects
and focuses a fast moving electron beam along tungsten target ring
in the gantry. Since the detector is also in the form of ring, it permits
simultaneous acquisition of multiple image sections.
The images are obtained in 50 ms times and can produce fast frame
rate CT movies of the beating heart with minimum motion artifacts.
The advantage is the speed of data acquisition. The whole heart can
be acquired in 0.2 s. These scanners are useful in cardiac imaging,
pediatric and trauma patients. It can also be used as conventional
CT by averaging multiple images with repeat scans.
SIXTH GENERATION
Third/fourth generation + slip ring technology + helical motion = Sixth
generation (1990). Slip ring is a circular contact with sliding brushes
and allows the gantry to rotate continuously. It eliminates inertial limitations
at the end of each slice and has greater rotational velocities, with shorter
scan time. Helical CT scanners acquire data while the table is moving.
The total scan time required to image the patient can be much shorter,
excluding time required to translate the patient table.
It allows the use of less contrast agent and increases patient throughput.
In some instances, the entire scan can be done within a single breathhold
of the patient. Raw data from helical scans can be interpolated to approximate acquisition of planar reconstruction data. The speed of
the couch motion is very important; hence the term pitch is defined.
SEVENTH GENERATION
Seventh generation uses multidetector array (MDA), the collimator spacing
is wider and more X-rays are used in producing image data. Opening
up the collimator in a single array scanner increases the slice thickness,
but reduces spatial resolution in the slice thickness dimension. Hence,
slice thickness is controlled by detector size, not by the collimator.
A 4 contiguous 5 mm detector array gives 20 mm collimator spacing.
The number of X-rays detected is 4 times higher than that of single
array of 5 mm. Further, 10 mm,15 mm, 20 mm slices may be obtained
from the same acquisition. Though it offers flexibility of CT acquisition
protocol, the number of parameters have increased. It has better efficiency
for patient imaging and detector pitch needs to be defined.
No comments:
Post a Comment