Basic Principles of Radiation Protection
Responsibility for Radiation Protection
A. Radiographer is primarily responsible for protecting the
patient from unnecessary exposure
1. Best accomplished by avoiding repeat exposures
2. Should use smallest amount of radiation that produces
a diagnostic radiographic image
B. Radiologist and referring physician assume shared
responsibility for radiation safety of the patient
1. Best accomplished by consultation
2. Should not order unnecessary exams
C. Safe use of radiation in diagnostic imaging to determine
the extent of disease or injury should outweigh the risk
involved from the exposure
Patient Exposure and Protection
The heart of radiation protection for the patient lies in the concept
of ALARA. It is primarily the radiographer’s responsibility
to see that ALARA is in practice so that patients are properly
protected. Taking adequate histories, communicating clearly,
using proper immobilization, and conducting radiographic and
fluoroscopic exams calmly and professionally add to the level of
safety your patients will encounter while under your care. This
section reviews the many technical aspects of the radiographer’s
practice that contribute to ALARA for the patient.
Beam Limitation
Beam limitation protects the patient by limiting the area of
the body and the volume of tissue being irradiated
A.Collimator
1. Variable aperture device
2. Contains two sets of lead shutters placed at right
angles to each other
3. Higher set of lead shutters is placed near the x-ray
tube window to absorb off-stem (off-focus) radiation
4. Lower set of lead shutters is placed near the bottom of
the collimator box to restrict the beam further as it exits
5. Accuracy of the collimator is subject to strict quality
control standards
6. Collimation should be no larger than the size of the
image receptor being used
7. Collimators that automatically restrict the beam to
the size of the image receptor have a feature called
positive beam limitation (PBL), also called automatic
collimation
8. PBL responds when an image receptor is placed in
the tray containing sensors that measure its size
B. Cylinder cones
1. Metal cylinders that attach to the bottom of the collimator
2. Used to restrict the beam tightly to a small circle
3. Diameter of the far end of the cone determines field
size
4. Cones may be extended an additional 10 to 12 inches
by a telescoping action for even tighter restriction of
the beam
5. Cones may be used for exam of the os calcis, various
skull projections, and cone-down views of vertebral
bodies
6. Use of cones results in a restriction of the x-ray beam
by cutting out a major portion of the beam
7. When cones are used, mAs must always be increased
to make up for the rays attenuated by the cone
8. Cylinder cones do not work by focusing the x-ray
beam down the cone; x-rays cannot be focused
C. Aperture diaphragm
1. Flat piece of lead with a circle or square opening in
the middle
2. Placed as close to the x-ray tube window as possible
3. Has no moving parts
Filtration
A filter is placed in the x-ray beam to remove long-wavelength
(low-energy) x-rays. Low-energy x-rays contribute
nothing to the diagnostic image but increase patient
dose through the photoelectric effect. As low-energy rays
are removed, the beam becomes “harder” (predominantly
short-wavelength, high-energy). This process of removing
low-energy rays results in a lower patient dose.
A. Two types of filtration: inherent and added
B. Inherent filtration
1. Glass envelope of the x-ray tube
2. Insulating oil around the tube
3. Diagonal mirror used for positioning light
C. Added filtration
1. Aluminum sheets placed in the path of the beam near
the x-ray tube window
2. Mirror placed in the collimator head
D. Total filtration
1. Equals inherent plus added filtration
2. Must equal at least 2.5-mm aluminum equivalent for
x-ray tubes operating at greater than 70 kVp
E. Half-value layer: Amount of filtration that reduces
the intensity of the x-ray beam to half of its original
value—measured at least annually by a qualified radiation
physicist
F. The radiographer never adjusts added tube filtration; if
it is suspected that the filtration has been altered, the
x-ray tube must not be used until checked by a radiation
physicist
Gonadal Shields
Gonadal shields are used to protect gonads from unnecessary
radiation exposure. They should be used whenever they
do not obstruct the area of clinical interest.
A. Gonadal shielding may reduce female gonad dose by up
to 50%
B. Gonadal shielding may reduce male gonad dose by up to
95%
C. Proper collimation may also greatly reduce gonadal dose
and should be used in conjunction with gonadal shields
D. Most commonly used gonadal shields
1. Flat contact shield: Flat piece of lead or a lead apron
placed over the gonads
2. Shadow shield: Suspended from the x-ray tube housing
and placed in the x-ray beam light field; requires
no contact with the patient; especially useful during
procedures requiring sterile technique
Exposure Factors
Exposure technique determines the quantity and quality of
x-rays striking the patient
A. Use optimal kVp for the part being radiographed
B. Use the lowest possible mAs to reduce the amount of
radiation striking the patient
C. The part being imaged should be measured with calipers
D. A reliable technique chart should be consulted to determine
the proper exposure factors to use
E. Use of automatic exposure controls (AEC) reduces the
number of repeat exposures
Grids
A. Result in an increase in patient dose because increased
mAs are required
B. Use appropriate type and ratio of grid for part being
imaged and exam being performed
Repeat Exposures
A. Always result in an increase in radiation dose to the
patient
B. Must be kept to a minimum
C. Should be tracked via a departmental repeat exposure
analysis
D. Reasons for repeat exposures should be documented
E. In-service education for areas of frequent repeat exposures
should be conducted by qualified radiographers or
radiologists
Technical Standards for Patient Protection
A. Minimum source-to-skin distance for portable radiography:
at least 12 inches
B. Fluoroscopy
1. Use of intermittent fluoroscopy (as opposed to a
constant beam-on condition)
2. Tight collimation of the beam
3. High kVp
4. Source-to-tabletop distance for fixed fluoroscopes:
not less than 15 inches
5. Source-to-tabletop distance for portable fluoroscopes:
not less than 12 inches
6. Proper filtration of the beam
7. Fluoroscopy timer that sounds alarm after 5 minutes
(300 seconds) of beam-on time
8. Fluoroscopy timer should not be reset before alarm
goes off; fluoroscopist must be made aware of time
that the patient and other persons in the room were
exposed
9. Exposure switch must be dead-man type
10. Limit dose at the tabletop to no more than 100 mGya
per minute
11. Limit use of high-level-control fluoroscopy (HLCF)
during interventional procedures to no more than
200 mGya per minute
12. Long exposure times (>30 minutes) can lead to skin
effects; the patient should be informed of the possibility
of skin injury (e.g., erythema, epilation)
13. Fluoroscopy times should be recorded
14. Pulsed fluoroscopy or low-dose modes should be
used when possible to achieve ALARA objectives
15. Personnel during fluoroscopy should stand on the
image intensifier side of the C-arm during lateral or
oblique projections
16. When possible, use the C-arm with the x-ray tube
below the patient for anteroposterior and posteroanterior
projections
17. During fluoroscopy, the radiographer should always
wear the radiation dosimeter outside of the lead
apron at the level of the collar
18. Avoid use of electronic magnification modes on
the image intensifier when possible because this
increases the dose to the patient and other persons
in the room
19. Air kerma
a. Provides an approximate skin dose where the
x-ray beam is entering the patient
b. Is displayed during a fluoroscopic procedure
c. Monitoring air kerma is a way to keep dose to
the patient lower, especially during prolonged
fluoroscopic procedures
d. Expressed as Gya (a indicates energy deposited in
a mass of air)
20. Dose area product (DAP)—the total of air kerma
striking the surface of the patient
a. May be read on the DAP meter on the fluoroscopic
monitor
b. Expressed as mGy-cm2
21. Last image hold (LIH)
a. Provides for keeping the most recently acquired fluoroscopic
image displayed on the monitor without
continued radiation exposure to the patient
b. Using last image hold reduces patient exposure
by allowing the fluoroscopist to evaluate an
image without continuing to expose the patient
22. Automatic brightness control (ABC) or automatic
exposure rate control (AERC)
a. kVp and mA are automatically adjusted during
fluoroscopy, which keeps image brightness level
b. The ABC or AERC delivers to the image receptor
only the exposure that is required to maintain
the necessary image quality
Cardinal Principles of Radiation Protection
A. Time: Amount of exposure is directly proportional to
duration of exposure
B. Distance: Most effective protection from ionizing radiation
1. Dose is governed by the inverse square law
2. The greater the distance from the radiation, the lower
the dose
3. Dose varies inversely according to the square of the
distance; for example, if the exposure is 5 mGya at a
distance of 3 feet, stepping back to a distance of 6 feet
causes the exposure to decrease to 1.25 mGya
4. The inverse square law should always be used during
fluoroscopy in which close contact with the patient
is not required and during mobile radiography and
fluoroscopy
C. Shielding: Lead-equivalent shielding absorbs most of the
energy of the scatter radiation
1. A lead apron of at least 0.25-mm lead equivalent must
be worn (0.5-mm lead equivalent should be worn)
during exposure to scatter radiation; a thyroid shield
of at least 0.5-mm lead equivalent should be worn for
fluoroscopy
2. The radiographer should never be exposed to the primary
beam
3. If exposure of the radiographer to the primary beam
is unavoidable, the exam should not be performed
4. Family members of the patient, nonradiology employees,
or radiology personnel not routinely exposed
should be the first choices to assist with immobilization
of the patient for an exam when all other types
of immobilization have proved inadequate
5. The radiographer should be the last person chosen to
assist with immobilization during an exposure
6. Radiographers and student radiographers should not
be viewed as quick and easy-to-use immobilization devices
7. Thickness recommendations for lead devices have
been determined by NCRP Report #102
Source of Radiation Exposure to Radiographer
A. Source of radiation exposure to the radiographer is scatter
radiation produced by Compton interactions in the
patient
B. Greatest exposure to the radiographer occurs during
fluoroscopy, portable radiography, and surgical radiography
C. It is vital that lead aprons be worn when fluoroscopy and
mobile radiography or fluoroscopy is performed
D. Photons lose considerable energy after scattering
E. Scattered beam intensity is about 1⁄1000 the intensity of
the primary beam at a 90-degree angle at a distance of 1
m from the patient
F. Beam collimation helps reduce the incidence of Compton
interactions, resulting in decreased scatter from the
patient
G. The use of high-speed image receptors may further
reduce the amount of scatter produced because of
decreased quantity of radiation needed for exposure
No comments:
Post a Comment